Cite as: Archiv EuroMedica. 2026. 16; 5. DOI 10.35630/2026/16/Iss.5.25
Temporary anchorage devices (TADs) provide skeletal anchorage in orthodontics, but reported complications and success and failure rates vary across studies.
To summarize evidence on TAD complications, success and failure rates, and factors investigated in relation to these outcomes.
PubMed, Europe PMC and BASE were searched on 20 August 2026. Eligibility was restricted to English publications with freely accessible full texts. Two reviewers independently selected studies and extracted data. Thirty-seven studies were included and synthesized descriptively without meta-analysis.
Outcomes varied across device types, anatomical sites, loading protocols and follow-up periods. Failures frequently occurred during the first few months after insertion. Root contact was associated with failure in several studies, and peri-implant inflammation was associated with early instability in one prospective study. Postoperative pain and sinus penetration were also reported. Findings concerning screw dimensions, insertion procedures, loading time and patient characteristics were inconsistent or specific to particular settings. Greater primary stability did not necessarily correspond to better clinical survival. The predictive value of inflammatory and microbiological findings remained uncertain. Differences in outcome definitions and incomplete reporting limited comparisons.
The available evidence does not support a single protocol for TAD selection, placement or loading. Future prospective studies should distinguish primary stability from clinical success throughout orthodontic loading and use standardized outcome definitions and reporting methods.
Keywords: temporary anchorage devices; orthodontics; complications; failure rates; anchorage stability; scoping review.
Temporary anchorage devices (TADs), including orthodontic mini-implants, micro-implants, palatal implants and miniplates, are widely used in contemporary orthodontics to provide skeletal anchorage and support the control of complex tooth movements. Their clinical value has been supported by generally high reported success and survival rates, often reaching or exceeding 90% in several reviews, although these outcomes vary depending on the type of device, placement site, patient characteristics and orthodontic protocol [20, 19].
Despite their established role in orthodontic treatment, TADs may be associated with complications or failure events, such as reduced stability, loosening or loss of the device, soft-tissue inflammation, discomfort, mucosal overgrowth, or injury to adjacent anatomical structures. Previous reviews have contributed substantially to the understanding of TAD effectiveness and survival; however, they have also highlighted important areas requiring further clarification. In particular, the evidence base is characterized by heterogeneity in study design, terminology, outcome definitions and clinical protocols, which makes direct comparison between studies challenging [20]. Similarly, differences between device types have been reported, with palatal implants and miniplates showing high survival rates and miniscrews demonstrating more variable outcomes, suggesting that device-related and clinical-context factors remain relevant when interpreting success and failure rates [19].
Several potential risk factors for TAD success or failure have been described in the literature and may be grouped into patient-related, device-related, procedure-related and orthodontic treatment-related factors [20]. However, the strength of evidence for specific factors remains uneven. For example, insertion torque has been frequently discussed as a clinically relevant parameter, but existing evidence does not clearly support a specific torque range as being associated with higher success rates. This uncertainty has been attributed to limitations such as subjective definitions of success, variability in measurement methods, and the need for more objective and standardized stability assessments [21].
The present review therefore focuses on a descriptive synthesis of reported complications, success and failure rates, and factors associated with TAD outcomes across different clinical settings.
A scoping review was chosen to summarize evidence from studies with different designs, clinical settings and outcome measures.
The aim of this review was to summarize the available evidence on complications and success and failure rates of TADs used in orthodontic treatment.
The specific objectives were to:
The population of interest was human patients of any age (children, adolescents, and adults) undergoing orthodontic treatment in which temporary anchorage devices (TADs), including orthodontic mini-implants and micro-implants, were used as skeletal anchorage. Studies conducted on animals or in vitro were excluded. Studies in which mini-implants were used exclusively for non-orthodontic purposes (e.g., purely surgical or prosthetic applications without an orthodontic anchorage role) were also excluded.
The main concept of the review was complications and failure or success rates associated with TADs in orthodontics. Eligible studies were required to report clinical outcomes related to adverse events and/or device stability, such as loss of stability, loosening or loss of the implant, soft tissue inflammation or infection, pain or discomfort, damage to adjacent anatomical structures, and overall success or failure rates. Studies that focused exclusively on biomechanical effectiveness or tooth movement outcomes without any data on complications or failure/success rates were excluded, as were purely technical or narrative articles that did not present clinical data.
The context comprised orthodontic treatment carried out in any clinical setting, including private practices, university clinics, and hospital departments, and in any anatomical location of TAD placement (maxilla, mandible, palate, etc.). Reports describing the use of similar devices solely in non-orthodontic contexts were not considered.
Regarding the types of evidence sources, the review included primary clinical studies such as randomized and non-randomized clinical trials, prospective and retrospective cohort studies, case-control studies, and larger case series (involving at least 8 patients or 20 TADs; the minimum sample-size threshold was defined before screening and applied consistently during study selection). Systematic reviews and other secondary studies were used to identify additional primary studies and to map existing evidence, but they were not considered the primary unit of analysis. Single case reports, very small case series, editorials, letters, and conference abstracts without sufficient clinical data or without an accessible full text were excluded.
Only articles published in English were considered. No lower date limit was applied; studies published up to and including 20 August 2026 were eligible. Furthermore, only studies with freely accessible full-text articles (open access or otherwise available without a paywall) were included in the review. The eligibility criteria are summarized in Table 1.
Table 1. Eligibility criteria
| Category | Inclusion | Exclusion |
| Population | Human patients of any age (children, adolescents, adults) undergoing orthodontic treatment with temporary anchorage devices (TADs), including orthodontic mini‑implants and micro‑implants, used as skeletal anchorage. | Animal studies; in vitro studies; studies where mini‑implants are used exclusively for non‑orthodontic purposes (e.g. purely surgical or prosthetic applications without an orthodontic anchorage role). |
| Outcomes | Studies reporting clinical outcomes related to complications and/or success/failure of TADs, e.g. loss of stability, loosening or loss of the implant, soft‑tissue inflammation or infection, pain/discomfort, damage to adjacent structures, overall success/failure rates. | Studies focusing exclusively on biomechanical effectiveness or tooth movement outcomes without any data on complications or success/failure; purely technical or narrative articles without clinical data. |
| Context (clinical setting) | Orthodontic treatment in any clinical setting: private practices, university clinics, hospital departments; any anatomical location of TAD placement (maxilla, mandible, palate, etc.). | Reports describing use of similar devices solely in non‑orthodontic contexts. |
| Study design / evidence | Primary clinical studies: randomized and non‑randomized clinical trials, prospective and retrospective cohort studies, case–control studies, and larger case series (≥8 patients or ≥20 TADs; final threshold predefined in protocol). | Single case reports; very small case series (below predefined threshold); editorials; letters; conference abstracts without sufficient clinical data or without accessible full text. Systematic reviews not used as primary units of analysis. |
| Language | Articles published in English. | Non‑English publications. |
| Time frame | Publications up to and including 20 August 2026. | Studies published after 20 August 2026. |
| Information sources | Studies identified via systematic search in PubMed (MEDLINE), Europe PMC and the Bielefeld Academic Search Engine (BASE); additional eligible studies from reference lists of included articles and relevant reviews. | Studies not retrievable from these sources and not meeting other inclusion criteria. |
The primary information sources for this scoping review were electronic bibliographic databases. A systematic search was conducted in PubMed (MEDLINE), Europe PMC and Bielefeld Academic Search Engine (BASE) to identify relevant clinical studies reporting complications and/or failure or success rates of temporary anchorage devices used in orthodontics. In addition, the reference lists of included articles and relevant reviews were screened to identify any further eligible studies that were not retrieved by the electronic database searches. Only studies with freely accessible full texts in English were considered.
A structured electronic search was conducted in PubMed (MEDLINE), Europe PMC, and the Bielefeld Academic Search Engine (BASE) using a universal query combining free-text terms related to temporary anchorage devices in orthodontics and their complications or failure/success outcomes. The same universal query was applied across all databases:
("temporary anchorage device" OR "temporary anchorage devices" OR "orthodontic mini implant" OR "orthodontic mini implants" OR miniscrew OR "mini screw" OR "micro implant" OR "skeletal anchorage" OR TAD) AND orthodontic AND (failure OR complication OR infection OR "success rate" OR "survival rate" OR stability OR loosening OR dislodgement OR "root contact" OR "peri-implant inflammation" OR "peri implant inflammation" OR "loading protocol" OR "CBCT planning")
The electronic searches of PubMed, Europe PMC, and BASE were conducted on 20 August 2026.
All records identified through the database search were imported into the Rayyan screening software, and duplicates were removed (T.H.). The remaining records were screened in two stages using Rayyan, a web-based systematic review management platform. Before formal screening, the reviewers completed a calibration exercise to ensure consistent application of the eligibility criteria. First, titles and abstracts were independently reviewed in a blinded manner to identify potentially relevant studies (K.R. and W.K.). Subsequently, the full texts of the selected articles were independently assessed for eligibility against the predefined inclusion and exclusion criteria (K.R. and W.K.). Any disagreements were resolved through discussion and consensus. Reasons for exclusion at the full-text stage were recorded. The study selection process was presented in a PRISMA-ScR flow diagram (Figure 1).
Data from the included studies were charted using a standardized data charting form developed by the research team. The form was pilot tested on a small sample of studies and refined as necessary. Two reviewers independently charted data from the selected articles using tables prepared in Google Sheets to ensure accuracy and consistency (K.R. and W.K.). Any discrepancies were resolved through discussion and consensus between the reviewers.
The following data items were collected from each included study when reported: (1) general study characteristics (first author, year of publication, country, study design, clinical setting, study period); (2) population characteristics (number of participants and/or number of TADs, age, sex distribution, relevant patient‑related factors such as oral hygiene, smoking status or systemic conditions, where available); (3) details of the temporary anchorage devices and their use (type of device, dimensions, manufacturer if reported, number of devices per patient, anatomical placement site, side/jaw, indication for use, loading protocol and timing, duration of use, operator characteristics); (4) outcomes related to complications and adverse events (type and number of events such as loss of stability, loosening, complete loss of the device, soft‑tissue inflammation or infection, mucosal overgrowth, pain or discomfort, damage to adjacent roots or other anatomical structures, and any other local or systemic complications); (5) outcomes related to success and failure (definitions of success/failure adopted in the study, number and proportion of successful and failed TADs, timing of failure, reported success or survival rates per device and/or per patient, follow‑up duration); (6) factors examined as potential determinants of complications or failure/success (patient‑, device‑ and procedure‑related variables, including results of any statistical analyses, if reported); and (7) other relevant methodological or reporting details (e.g. inclusion/exclusion criteria, method of outcome assessment, handling of missing data, funding or conflict‑of‑interest statements).
The results were synthesized descriptively in line with the scoping review methodology. Extracted data were first organized in tabular form to provide an overview of study characteristics, populations, TAD types and placement sites, as well as reported complications and success/failure outcomes. Studies were then grouped and compared according to key dimensions such as study design, anatomical location, type of device, clinical indication and patient population, where feasible. Frequencies and ranges of reported complications and success/failure rates were summarized narratively, and simple descriptive statistics (e.g. counts, proportions and ranges) were used where appropriate. Particular attention was paid to how success, failure and specific complications are defined and measured across studies, as well as to any reported associations with patient-, device- or procedure-related factors. No formal meta-analysis or quantitative pooling of data was performed, as the primary aim was to map and describe the breadth and nature of the existing evidence rather than to estimate pooled effect sizes.
The electronic searches in PubMed, Europe PMC, and the Bielefeld Academic Search Engine (BASE) yielded a total of 1464 records. After the removal of 582 duplicates, 882 records remained for screening. Titles and abstracts were screened against the eligibility criteria, resulting in the exclusion of 836 clearly irrelevant records (e.g., animal or in vitro studies, non-orthodontic use of mini-implants, non-clinical articles, non-English publications, and studies without accessible full text). The full texts of 46 potentially relevant articles were then assessed in detail, of which 37 met all inclusion criteria and were included in the scoping review. The remaining 9 full-text articles were excluded for predefined reasons (e.g., absence of data on complications or failure/success rates, sample size below the threshold for inclusion, conference abstract only, case report, or very small case series) and were listed in Table 2 with the specific reason for exclusion. The study selection process, including the number of records at each stage and the reasons for exclusion at the full-text level, was presented in a PRISMA-ScR flow diagram. Screening of the reference lists of included studies and relevant reviews identified no additional eligible studies.
Table 2. Full-text articles excluded after eligibility assessment and reasons for exclusion
| Title | Author, year | ID | Reason for exclusion |
| Clinical transfer accuracy of pressure-moulded versus 3D-printed drilling guides for orthodontic mini-implants in the anterior palate: a randomized prospective clinical study | Möhlhenrich SC et al., 2022 | https://doi.org/10.1038/s41598-026-50346-z | absence of data on complications or failure/success rates |
| Assessment of interradicular areas and cortical bone thickness for orthodontic mini-implant placement: a cone-beam computed tomography study | Gupta SP et al., 2025 | https://doi.org/10.1186/s12903-025-06251-4 | absence of data on complications or failure/success rates |
| Radiographic Evaluation of Implant Stability and Osseointegration in Adult Orthodontic Patients | Rostamzadeh S et al., 2025 | https://doi.org/10.31661/gmj.v14i.3799 | review article |
| Comparative analysis of two navigation techniques based on augmented reality technology for the orthodontic mini‑implants placement | Deglow ER et al., 2023 | https://doi.org/10.1186/s12903-023-03261-y | in vitro study |
| Comparative evaluation of orthodontic mini-implants hygiene protocols on subgingival bacterial load | Freitas AOA et al., 2024 | https://doi.org/10.1590/2177-6709.30.1.e2524108.oar | absence of data on complications or failure/success rates |
| Correlation between cortical bone thickness at mini-implant insertion sites and age of patient | Centeno ACT et al., 2020 | https://doi.org/10.1590/2177-6709.27.1.e222098.oar | absence of data on complications or failure/success rates |
| Insertion of orthodontic temporary anchorage devices with free gingival grafting for phenotype modification of the peri-implant mucosa | Maldonado Molina O, 2023 | https://doi.org/10.1016/j.jobcr.2023.09.005 | small case report |
| Reasons for mini-implants failure: choosing installation site should be valued! | Consolaro A, Romano FL, 2014 | https://doi.org/10.1590/2176-9451.19.2.018-024.oin | review article |
| Microbial Colonization in Orthodontic Mini-Implants | Freitas AOA et al., 2012 | ISSN 0103-6440 | absence of data on complications or failure/success rates |
Figure 1. PRISMA-ScR flow diagram of study selection
The included studies span randomized trials, prospective and retrospective cohorts, and a case–control design, conducted in clinical orthodontic settings and covering common TAD types, sizes, and placement sites (buccal alveolar, palatal, extra‑alveolar). They report success/survival, timing of failures, and selected complications, and examine patient-, device-, anatomic-, and procedural risk factors. Detailed study characteristics and outcomes are presented in Table 3.
Table 3. Characteristics and main findings of included studies
| Author, year | Study design | Participants (n) | TADs (n), type & site | Definition of success / failure | Follow‑up / loading | Main quantitative outcomes | Main factors examined |
| Zhao et al., 2023 | Case–control study | 47 patients (for metagenomics subset; smaller subsets for 16S and qPCR) | 29 TADs for 16S (15 failed, 14 successful); 135 TADs for metagenomics (62 failed, 73 successful); 34 TADs for qRT‑PCR (18 failed, 16 successful); self‑drilling Ti TADs, 7–8 mm, various alveolar sites | Failed: TADs showing severe mobility with clinical signs of inflammation and unable to serve as anchorage before end of treatment; Successful: TADs stable until end of treatment with no signs of inflammation or infection | Variable in‑mouth time; all TADs loaded 1 month after insertion; failed TADs had significantly shorter retention time (p < 0.001) | Beta-diversity differed between failed and successful TADs (PERMANOVA p < 0.05). Failed TADs were enriched in Prevotella intermedia, Eikenella corrodens, Parvimonas spp., Neisseria elongata, and Catonella morbi. Prevotella intermedia had ROC AUC 0.861 and higher load in failed TADs (qRT‑PCR p = 0.0048). | Microbiome differences (higher bacterial load, especially P. intermedia; motility/metabolic pathways) - key clinical modifiers: age, oral hygiene/periodontal status, retention time. |
| Yao et al., 2014 | Large retrospective clinical cohort | 220 patients (66 males, 154 females; mean age 29.3 yrs) | 727 mini‑implants total; 643 first‑time implants and 84 re‑implants. Three types: 159 titanium miniplates, 388 pre‑drilled Ti miniscrews (L‑miniscrew), 180 self‑drilling stainless‑steel miniscrews (J‑miniscrew); sites: upper/lower arch; buccal, palatal, edentulous alveolar ridge; anterior/posterior to 2nd premolars | Failure: significant loosening/mobility that could not withstand orthodontic loading (necessitating replacement or abandonment); Success: remained stable and functional under orthodontic load for intended anchorage period | Retrospective over 5‑year period (2007–2012); loading typically after healing, but early (≤30 d) and late (>30 d) loading both present; up to 4 implantations per patient | First implantation failure rates: 5.4% (miniplates), 19.1% (Ti L‑miniscrews), 25.4% (SS J‑miniscrews); failure rates increased with each re‑implantation, especially for SS miniscrews (73.3% at 2nd implantation, 66.7% at 3rd). GEE: OR for failure ≈5.1 (L‑miniscrew) and ≈5.3 (J‑miniscrew) vs miniplate; age <35 yrs OR ≈11.7 vs ≥35; late loading (>30 d) OR ≈2.0 vs ≤30 d; edentulous ridge site OR ≈12.9 vs buccal/palatal alveolar process (p≈0.055) | Mini‑implant type (plate vs Ti vs SS), number of implantations (first vs re‑implant), patient age, sex, malocclusion class, facial divergence (SN‑MP), arch (upper/lower), site (buccal/palatal/ridge; anterior/posterior), clinical bone density (D2–D4), soft‑tissue type (attached gingiva, mucogingival junction, movable mucosa), soft‑tissue inflammation (none/mild/moderate/severe), healing time (≤30 vs >30 days before loading), dentofacial cephalometric variables (SNA, SNB, ANB, SN‑MP, etc.) |
| Ravi et al., 2022 | Single‑centre randomized split‑mouth clinical trial | 9 recruited, 8 completed (all female, 23–29 yrs) | 36 mini‑implants placed, 32 analyzed for failure; self‑drilling tapered Ti miniscrews, 2.0×8 mm (A1, Bioray); buccal interradicular region between 2nd premolar and 1st molar in all four quadrants; one surface‑treated and one non‑treated per patient | Success: mini‑implant with at most minimal mobility that could resist orthodontic load and remained in bone until completion of space closure; Failure: loosening during treatment such that it could not resist orthodontic force and required relocation/removal | Healing period 4 weeks, then loaded with 150 g per side using NiTi closed‑coil springs for en‑masse retraction on 0.019×0.025‑in SS archwires; removal at end of space closure, when removal torque recorded | Mean insertion torque: 17.9 ± 5.6 Ncm (surface‑treated) vs 16.4 ± 9.0 Ncm (non‑treated), NS (p=0.674); mean removal torque: 8.1 ± 2.9 Ncm (surface‑treated) vs 3.3 ± 1.9 Ncm (non‑treated), p=0.009; of 7 failed implants, 71.4% were non‑treated and 28.6% surface‑treated (difference in failure rates not statistically significant, χ² p=0.394) | Surface treatment (sandblasting + acid‑etching vs machined); primary stability (maximum insertion torque); secondary stability (maximum removal torque); failure rate distribution between surface‑treated vs non‑treated miniscrews |
| Umeh et al., 2023 | Retrospective clinical survival study | 63 patients (M:F ≈ 1:2; mean age 28.0 ± 11.5 yrs) | 90 TADs (Speed Dental, Korea), diameter 1.6 mm, lengths 8 mm (n=70) and 10 mm (n=20); placed in buccal, palatal and labial segments of upper and lower arches | Failure: TAD showing mobility or complete dislodgement within 6‑month observation; Success: TAD present and stable (no mobility/dislodgement) at 6 months | 6 months; all TADs immediately loaded after placement; 4‑weekly reviews | Overall survival 88.9% (80/90); 10/90 failures (11.1%); of these, 8 (8.9%) occurred in first month, 1 (1.1%) in 2nd month, 1 (1.1%) in 3rd month; no failures between 4–6 months (p=0.001 for association between failure and time). Highest failure rate in age 11–20 group (15.6%), 100% survival in 21–30 yrs, but age effect NS. No failures in palatal or labial sites (all 10 failures in buccal sites); no failures with 10‑mm screws (all 10 failures with 8‑mm screws), but differences by length/site/jaw not statistically significant. | Demographics: age group, gender; device factors: length (8 vs 10 mm); site: buccal vs palatal vs labial; side (left vs right); arch (maxilla vs mandible); time since placement (months 1–6) and its association with failure. |
| Jeong et al., 2015 | Retrospective clinical time‑to‑event analysis | 134 patients (mean age 22.08 ± 7.52 yrs) | 331 self‑drilling miniscrews (Miangan, Korea), 1.2 mm diameter, 7.0 mm length; buccal alveolar bone between first premolar and second molar in maxilla and mandible | Failure group (FG): miniscrews that failed (loosened/removed); Non‑failure group (NFG): miniscrews that remained stable. FG subdivided into failed before loading (FGB) and failed after loading (FGA). Failure implicitly defined as loss of stability requiring removal; success as continued use throughout observation under orthodontic load. | Observation up to 52 weeks; miniscrews inserted after leveling/alignment; loading applied to NFG and FGA; loading time after insertion varied (1–40 weeks) | 57/331 failures (17.22%); 28 before loading, 29 after loading. About 75.44% of all failures occurred within 16 weeks after insertion. Distribution of failures after insertion: 15.79%, 36.84%, 12.28%, 10.53% at 4, 8, 12, and 16 weeks. Failure rate according to loading time after insertion (FGA/NFG): 13.56%, 8.97%, 11.32%, 5.00% at 4, 8, 12, and 16 weeks; miniscrews loaded during first week after insertion had 42.5% failure. After loading, 58.62% of failures occurred between 4–12 weeks; failures after loading observed mainly within first 12–13 weeks, with occasional events up to 35–44 weeks. Authors conclude stability is typically acquired 12–16 weeks post‑insertion and immediate loading can cause failure. | Temporal pattern of failure after insertion, temporal pattern after loading, failure rate as a function of loading time after insertion. No analysis of demographic or anatomical predictors in this paper; focus is exclusively on timing of failure and implications for recommended loading delay (12–16 weeks). |
| Gurdán & Szalma, 2018 | Retrospective clinical cohort study | 47 patients | 59 self‑drilling miniscrews (Jeil Dual Top Anchor System, 1.6 × 8 mm); maxilla (91.5%) and mandible (8.5%); sites: palate (54.2%), buccal fold (40.7%), ascending ramus (5.1%); anterior (16.9%) and lateral (83.1%) regions | Success: provided “perfect skeletal anchorage” throughout the treatment period, without any sign of mobility; Failure: any screw showing mobility/loosening (with or without symptoms), peri‑implant infection, screw fracture, or neighbouring tooth injury | Mean loading time 8.1 ± 3.3 months; all screws immediately loaded (force <150 g), with some used simultaneously with surgical exposure of impacted teeth | Overall success 89.8% (53/59); overall failure 10.2% (6/59). Peri‑implant inflammation in 6/59 (10.2%): palate 6.25%, buccal fold 12.5%, ascending ramus 33.3% (small n). Screw mobility 6/59 (10.2%): palate 3.1% vs buccal fold 20.8% (p=0.034); 0% in ascending ramus. By treatment type: loosening significantly more frequent in intrusion (2/8, 25.0%) than in extrusion of impacted teeth (1/32, 3.1%) (p=0.036); distalization 3/16 (18.8%); uprighting 0/3 (0%). One screw fracture (33.3% of ramus cases). No radiographic root injuries detected. | Demographics: age, sex; jaw (maxilla vs mandible); anatomic localization (palate vs buccal fold vs ascending ramus); jaw region (anterior vs lateral); type of orthodontic movement (extrusion of impacted tooth, intrusion, distalization, uprighting); incidence of screw mobility and peri‑implant inflammation by site and by type of movement. |
| Tsai et al., 2016 | Prospective clinical cohort with Kaplan–Meier survival analysis and Cox regression | 139 patients (114 females, 25 males; mean age 25.7 ± 7.5 yrs, range 12–56) | 254 mini‑implants (MIs): stainless steel and Ti‑alloy; 2.0 mm diameter (all), lengths 2×12, 2×11, 2×10, 2×9, 2×8, 1.5×9 mm; sites: maxilla 83.9%, mandible 16.1%; anterior 7.1%, posterior 92.9%; left 40.6%, right 39.8%; used for retraction, intrusion, uprighting, reinforced anchorage, BAMP | Failure: MI removed due to loosening, pain, infection, or pathologic soft‑tissue change; Success: remained functional (no such events) through observation; censored if stable at 1 year or removed after successful completion of movement | 1‑year observation; early loading ≤3 weeks vs late loading >3 weeks; forces mostly via elastomeric chains, occasionally inter‑arch elastics or ligatures | Overall success was 85.8% (218/254); 1‑year cumulative survival 81.6%. Younger patients had longer survival — 20–30 yrs fared better than >30 yrs (log‑rank p=0.011). In the Cox model, each additional year over 30 increased hazard (HR 1.05 per year, p=0.008), ~2.66-fold higher risk over 20 years. Longer miniscrews (10–12 mm) had better 1‑year survival than shorter ones (8–9 mm) (p=0.004); HR ≈ 2.62 for short vs long (95% CI 1.25–5.50). Sex, malocclusion, facial divergence, jaw, position, side, material, treatment purpose, force mode, and healing time showed no significant effect. | Patient‑related: age (<20, 20–30, >30 yrs), sex, malocclusion class (I/II/III), facial divergence (high/avg/low), operator. Implant‑related: material (SS vs Ti‑alloy), jaw, position (anterior/posterior), side, purpose (retraction, intrusion, uprighting, etc.), size (diameter fixed; length 8–12 mm, 1.5×9), loading time (≤3 vs >3 weeks), mode of force application. Primary focus on age and MI length as predictors of survival. |
| Yamaguchi et al., 2024 | Retrospective CBCT‑based clinical cohort with multivariable logistic regression | 60 patients (11 males, 49 females; mean age 21.0 ± 7.0 yrs) | 101 TADs (Vector TAS, Ormco), 1.4 mm diameter, 8.0 mm length; all placed in buccal alveolar bone between maxillary 2nd premolar and 1st molar (right: 51, left: 50) | Success: TAD usable as anchorage for ≥6 months after implantation; Failure: loss of anchorage (loosening/dislodgement) within 6 months | 6‑month observation; all devices used clinically as anchorage; loading protocol not detailed beyond routine orthodontic use | Overall failure rate was 15.8% (16/101). In bivariate tests, only vertical skeletal pattern (higher FMA) differed between failures and successes (p = 0.024); left–right, sex, age, ANB, bone density, and contact length showed no difference. In multivariable logistic regression, root contact was the sole independent predictor of failure (OR 28.25; 95% CI 5.29–151.04; p = 0.001). Twelve of 25 TADs with root contact failed versus 4 of 76 without (p < 0.001). Tooth type in contact (2nd premolar vs 1st molar) was not significant (p = 0.202). | Sex, age; sagittal skeletal pattern (ANB), vertical skeletal pattern (FMA); CBCT‑derived cortical and cancellous bone density (predicted HU); cortical and cancellous bone contact length; presence/absence of root contact; tooth type contacted (2nd premolar vs 1st molar); maxillary sinus perforation; side (left vs right). Primary focus on CBCT‑based anatomic risk factors, especially root contact. |
| Andrucioli MC et al., 2019 | In vivo, observational comparative study | 13 patients (11–49 years old) | 18 mini‑implants (11 successful, 7 unsuccessful); Neodent miniscrews 1.6×7 or 9 mm; both jaws | Success: well-anchored mini-implants removed after completion of mechanics/treatment; Failure: mini-implants loosened before achieving desired tooth movement, removed earlier | Mean time in mouth: success 29.4 months, failure 7.6 months; all with primary stability and identical hygiene recommendations | RT-qPCR: mRNA expression (as % of GAPDH) of IL-1α, IL-6, IL-17, TNF-α, RANK, RANKL, OPG; no significant differences between groups for IL-1α, IL-17, TNF-α, RANK, RANKL, OPG (p>0.05); IL-6 significantly higher in the failed mini-implant group (p=0.0397); no difference in the RANKL/OPG ratio | Proinflammatory cytokines (IL-1α, IL-6, IL-17, TNF-α) and osteoclastogenesis markers (RANK, RANKL, OPG); association of IL-6 expression with loss of mini-implant stability |
| Haddad & Saadeh, 2019 | Retrospective observational | 260 patients (131 M, 129 F; mean age 23.45 y; 13–51.4 y) | 260 MIs: maxilla 209, mandible 51; AbsoAnchor 1.4×8 mm; Imtec 1.8×8 mm; sites: C–Pm1, Pm1–Pm2, Pm2–M1, M1–M2; placed 30–35° in attached gingiva | Success: stable until completion of required movement; Failure: instability observed | Immediate loading ~150 g; periapical orthoradial radiographs (DT, DC) | Success 88.1% (229/260). DC higher in success: 7.46±1.7 mm vs 3.43±0.81 (p<0.001). DT not different (p>0.05). Logistic regression: failure ↓ with age (Coef −0.345; p=0.013), inter‑premolar (p=0.028) and premolar–molar sites (p=0.045), and with increasing DC (Coef −3.595; p<0.001). | Distance to alveolar crest (DC), root proximity (DT), age, jaw, side, site, MI type; DC strongly linked to stability; RP not associated with failure. |
| Alsaeedi AF et al., 2024 | Split-mouth, double-blind RCT (1:1), single-operator | 24 patients (18–30 y) | 48 IZC TADs (Tusk Dental), lengths 12×2 mm and 14×2 mm; bilateral IZC between maxillary first and second molars; self-drilling; initial 90° then 55–70° apical angulation; CBCT verification | Success: maintained stability to achieve mechanics; Failure: removal due to loosening/inflammation/fracture or spontaneous loss | Immediate loading 227–397 g per side; stability checked up to 6 months | Primary stability higher with 14 mm (46.5 vs 44.3; p=0.015). Late stability at 2 mo: NS (37.5 vs 36.5; p=0.176). Sinus penetration: 62.5% (14 mm) vs 54.2% (12 mm), NS. Failure: 16.7% (14 mm) vs 20.8% (12 mm), NS. Pain day 1 higher with 14 mm; at 1 week mild pain 12.5% (14 mm) vs 4.1% (12 mm). | Mini-implant length (12 vs 14 mm), primary/late stability, sinus penetration (CBCT), pain (VAS), failure rate; recommendation: prefer shorter IZC TADs due to similar failure but less risk to adjacent structures. |
| Chen YC et al., 2025 | Split‑mouth randomized clinical trial; guide vs real‑time navigation; retrospective freehand negative control | 11 pts; 61 TADs total; RCT arms: 16 sites each; freehand: 29 sites | Temporary anchorage devices; maxilla/mandible; planned via CBCT; guide (3D‑printed) vs IRIS navigation; freehand from records | No immediate complications requiring removal; monitored for peri‑implantitis/loosening | Immediate post‑op CBCT; planning/registration times recorded | Platform center deviation (mean±SD): guide 1.17±0.77 mm vs navigation 2.99±1.75 mm (p=0.010); freehand 3.10±1.91 mm. Angular deviation: guide 12.43±8.32° vs navigation 12.49±4.48° (NS); freehand 15.64±5.94° (NS vs others). No immediate complications in any group. | Accuracy (platform, angular deviation), safety (complications), method: surgical guide vs real‑time navigation vs freehand; combined guide+navigation suggested trend to higher accuracy. |
| Uchida Y et al., 2021 | Observational cohort (retrospective) | 83 patients (13.3–59.5 y) | 161 OASs; self-tapping, 2.0 mm diameter, 6.0 mm length; mid‑palatal area (mesial and distal to maxillary first molar) | Success: withstood orthodontic force ≥6 months without loosening/detachment | Horizontal 2.0–4.0 N via palatal attachment; ≥6 months | Success rate: pre-drill 1.2 mm = 94.5% vs 1.5 mm = 83.0% (p<0.05). Mesial zone: 100.0% (1.2 mm) vs 77.3% (1.5 mm) (p<0.005). Distal zone: 89.2% vs 88.6% (NS). Torque (1.2 mm group): success 25.9 N·cm vs failure 19.2 N·cm (p<0.05). | Pre‑drilling diameter (1.2 vs 1.5 mm), placement zone (mesial/distal), placement torque, sex; smaller pilot (≈60% of screw Ø) and higher torque (≈20–30 N·cm) associated with greater stability, especially mesially. |
| Moon CH et al., 2010 | Retrospective observational cohort | 306 patients (10–55 y) | 778 OMIs; Dual‑Top, 1.6×8 mm, cylinder, self‑drilling; maxillary/mandibular posterior buccal (U1–U3, L1–L3) | Success: ≥10 months loading without pain/mobility or goal achieved | Load 150–200 g at 2–3 weeks | Overall success 79.0%; ~80% of failures within 4 months (32.5% in first month). Placement position differed in success (p<0.05) but ORs NS. Avg upper gonial angle success 84.2% vs low 75.7% vs high 71.2% (p<0.01). High FMA had lower odds (OR 0.167; p=0.035). Avg upper gonial angle higher odds (OR 1.932; p=0.019). | Clinical: sex, age, soft‑tissue, side, sagittal class, ALD (NS); placement position (rate diff only). Skeletal: vertical pattern (FMA, upper gonial) associated with success; other angles NS. |
| Martha et al., 2026 | Retrospective observational study | 60 patients and 121 immediately loaded TADs | 121 self-drilling TADs placed palatally, in the upper vestibule, or lower vestibule | Success: stability without mobility, pain, inflammation, or premature loss during treatment | Clinical follow-up; bone quality assessed using records, radiographs, and CBCT when available | Overall success: 89.3% (108/121). Palatal: 80.0%; upper vestibular: 93.6%; lower vestibular: 94.7%. Associations were not statistically significant | Bone quality; insertion site; age; sex; immediate loading; inflammation; oral hygiene; TAD design |
| Bettenhäuser-Hartung et al., 2026 | Retrospective observational study | 283 patients; 199 females and 84 males; mean age 27.6 ± 13.3 years | 958 self-drilling interradicular mini-screws, 10 mm long; 496 palatal and 462 buccal. Used for maxillary total arch distalization with completely customized lingual appliances | Success: sufficient stability until the screw completed its anchorage function. Failure: spontaneous loss or removal due to mobility before completion of its intended function | Standardized insertion protocol with 1-mm pilot drilling, controlled angulation, and immediate loading with 150–200 cN. Mean loading period: 6.9 months | Overall premature failure: 1.5% (14/958). Failure was 2.6% palatally vs 0.2% buccally (p = 0.002). The highest failure rate occurred in patients aged >45 years (4.4%; p = 0.006*). Operator experience, sex, and quadrant were not significantly associated with failure | Buccolingual position; patient age; clinician experience; mesiodistal position; sex; quadrant; insertion protocol; immediate loading; mini-screw stability; treatment duration |
| Omer, 2018 | Prospective observational study | 47 orthodontic patients and 47 TADs | Bio-Tack mini-implants, 8–10 mm long and 1.4–1.8 mm in diameter; maxillary and mandibular anterior and posterior regions | Success: stability without mobility or significant inflammation. Failure: mobility, instability, or loss | Immediate loading with <200 g; follow-up at baseline, 1 day, 1 week, and 2 weeks | Overall success: 85.1% (40/47). Inflammation was associated with failure: success after 1 week was 92.9% without inflammation vs 33.3% with moderate inflammation (p = 0.020) | Inflammation; immediate loading; pain; age; sex; TAD dimensions; mucosal type; insertion site; oral hygiene |
| Yoshida et al., 2022 | Retrospective observational study | 197 orthodontic patients; 41 males and 156 females; mean age 20 years 2 months. A total of 467 TADs were evaluated | 467 self-drilling TADs of five types; diameter 1.3–1.6 mm and length 5–8 mm. Devices were placed in the attached gingiva on the buccal side of the molar region: 357 in the maxilla and 110 in the mandible | Failure: TAD mobility, detachment, or loss requiring removal. Failure was recorded within 750 days after implantation | All TADs were placed by one orthodontist under local anesthesia using a self-drilling technique. No orthodontic force was applied for at least 1 month, usually 2–3 months. A pilot hole was prepared in 82 cases (17.6%). Failure timing was analyzed in 30-day intervals | Overall failure rate: 15.6% (73/467). Failure was higher in males than females (23.9% vs. 13.6%; p = 0.024) and in the mandible than maxilla (28.2% vs. 11.8%; p < 0.001). Patients aged ≥30 years had the highest failure rate (29.8%; p = 0.029). Approximately 47.6% of maxillary failures occurred within 120 days, while 58.1% of mandibular failures occurred within 60 days. Early failure increased as FMA decreased, particularly in the maxilla (p = 0.005) and overall (p < 0.001). FMA was not significantly associated with failure after 120 days | Sex; age; jaw; implant side; FMA; ANB; overbite; overjet; time to failure; TAD type, diameter, and length; pilot drilling; and potential effects of cortical bone thickness, density, and insertion torque |
| Tseng et al., 2018 | Prospective observational study | 68 patients; 41 females and 27 males; mean age 27.7 ± 5.13 years; age range 18–37 years | 104 self-drilling stainless-steel miniscrew implants: 66 placed in the mandibular buccal shelf (BS; 2.0 × 12 mm) and 38 placed in mandibular interradicular (IR) sites (1.5 × 8 mm). | Success: maintenance of implant stability during orthodontic treatment, assessed by continued function and RF measurements. Failure: loss of the miniscrew during treatment; 13 implants failed | RF values were recorded at baseline (T0) and at 3, 6, 9, 12, and 15 weeks after placement (T1–T5). Orthodontic force of 150–200 g was applied after 3 weeks using elastomeric chains | Overall success 87.5% (91/104); failure 12.5% (13/104). Failures: 11 at week 15, 1 at week 9, 1 at week 12. Resonance frequency (RF) was higher at BS sites than IR sites during early healing. Adjusted mean RF fell from 11.0 kHz (T0) to 4.9 kHz (T5) at BS, and from 8.2 kHz to 3.7 kHz at IR. Implantation site and time both significantly affected RF (P = 0.003 and P < 0.0001); the site×time interaction was also significant (P < 0.0001). | Implantation site (buccal shelf vs interradicular), follow-up time, sex, patient age, insertion side, implant dimensions, bone quality and quantity, primary and secondary stability, healing-related changes, and timing of orthodontic loading |
| Ichinohe et al., 2019 | Observational clinical study | 25 patients; 50 mini-screws | Midpalatal mini-screws; two screws per patient | Success: retention for >6 months. Failure: spontaneous loss or removal because of mobility | Follow-up for at least 6 months | Success rate: 84% (42/50). Success was higher with cortical bone thickness ≥1.5 mm, insertion depth ≥4.5 mm, and screw–suture distance of 1.5–2.7 mm | Cortical bone thickness, insertion depth, screw–suture distance, soft-tissue thickness, nasal-cavity perforation, age, sex, and screw location |
| Ikenaka et al., 2022 | Retrospective observational study | 59 patients; 109 miniscrews | Titanium-alloy miniscrews (1.6 × 8.0 mm), placed between the maxillary second premolar and first molar on the buccal side | Failure: miniscrew mobility or loss during the first year | Orthodontic loading after a 1-month healing period; CBCT obtained immediately after placement | Overall failure rate: 14.7% (16/109). Root contact was the strongest risk factor (OR 13.50; 95% CI 3.11–58.55). Among screws with root contact, failure risk approximately doubled for each 1-mm increase in contact length (OR 2.07; 95% CI 1.11–3.84) | Root contact, root-contact length, age, sex, ANB, FMA, and implantation side |
| Azeem et al., 2019 | Retrospective pilot observational study | 40 patients; 60 mini-implants. Patients: 23 women and 17 men; mean age 20.1 ± 8.9 years | Orthodontic miniscrews measuring 1.3 or 1.5 mm in diameter and 8 or 10 mm in length, placed in the maxillary tuberosity region for maxillary molar distalization | Success: functional stability until the end of orthodontic treatment or intentional removal. Failure: detectable mobility or loosening during orthodontic treatment | Miniscrews were immediately loaded with 100–150 g using elastomeric chains or nickel-titanium coil springs. Mean force-application period was approximately 58 weeks | Overall failure rate: 26.3% (success rate: 73.7%). Operator experience was the only significant predictor: miniscrews placed by inexperienced operators had a 4.16-fold greater odds of failure (OR 4.16; 95% CI 2.071–6.321; P = 0.001). Age, sex, miniscrew diameter and length, force magnitude and delivery method, oral hygiene, implant side, and inflammation were not significantly associated with failure | Operator experience, patient age and sex, miniscrew diameter and length, orthodontic force magnitude, force-delivery method, oral hygiene, implant side, inflammation, and mandibular plane angle/vertical growth pattern |
| Geshay et al., 2021 | Prospective feasibility study | 26 adults; 82 miniscrew implants. Participants included 10 men and 16 women; mean age 27.4 years | 3 mm long and 1.7 mm wide miniscrew implants, placed in the buccal maxilla. Anterior sites were between the canine and first premolar; posterior sites were between the second premolar and first molar | Success: miniscrew remained clinically stable without mobility. Failure: any degree of mobility upon examination. Failures were classified as primary failures or incidental traumatic dislodgement | Adjacent miniscrews were immediately loaded with 100 g. Follow-up was performed at 1, 3, 5, and 8 weeks. Insertion and removal torque were also measured | Overall failure 32.9% (27/82); primary failure (excluding traumatic dislodgement) 23.6% (17/72). Anterior vs posterior failure: 35.7% vs 30.0%; primary failure: 30.1% vs 15.2%. Early learning effect: first 41 implants failed more than last 41 (46.3% vs 19.5%; P = 0.010). All primary failures occurred by day 42 (mostly weeks 2–4). Mean insertion torque ≈7.8 Ncm anterior, 7.4 Ncm posterior; mean removal torque ≈1.7 Ncm. Pain/discomfort minimal and short-lived (median week‑1 scores 2.2% and 5.5% of maximum). | Operator experience, anterior versus posterior placement site, implant side, immediate loading, insertion and removal torque, early implant movement/tipping, gingival thickness, timing of failure, traumatic dislodgement, and patient-reported pain and discomfort |
| Ashil et al., 2024 | Prospective comparative clinical study | 15 patients aged 13–26 years; 30 implantation sites. The patients were divided into a CBCT group and an RVG group, with 15 mini-implant sites per group | Orthodontic mini-implants placed in the buccal interradicular region between the maxillary or mandibular second premolar and first molar. In the CBCT group, placement sites were determined using CBCT. In the RVG group, placement was guided by two-dimensional digital radiography and a customized surgical guide | Success: accurate placement close to the predetermined optimal position without root contact. Failure/inaccuracy: deviation from the planned position, particularly vertical displacement or contact with an adjacent root. No implant survival or clinical stability endpoint was directly evaluated | CBCT scans were obtained before implantation and three months after implantation. The study assessed deviations in implant position, including vertical height (DMH), mesiodistal and transverse deviations (DMEP and DMT), and angular deviation (DMA) | The CBCT group showed significantly lower vertical deviation than the RVG group: mean DMH 0.0876 mm versus 0.546 mm (P = 0.01). Differences in DMEP (0.365 vs 0.623 mm; P = 0.132), DMT (0.575 vs 1.028 mm; P = 0.198), and DMA (4.5° vs 6.1°; P = 0.615) were not statistically significant. No mini-implants in the CBCT group contacted a root, whereas root contact occurred in 15% of implants placed in the RVG group. The RVG implants were positioned at a slightly lower height, closer to the archwire | Imaging modality, CBCT-based treatment planning, two-dimensional radiographic guidance, customized surgical guide, vertical placement deviation, mesiodistal and transverse deviation, angular deviation, maxillary versus mandibular site, and root contact |
| Yassaei et al., 2023 | Randomized, double-blind, split-mouth controlled clinical trial | The abstract reports 18 patients; the final statistical analysis included 11 patients. Participants were 12–15 years old with Class II division 1 malocclusion requiring extraction of the maxillary first premolars and anterior retraction | Maxillary orthodontic miniscrews, 1.6 × 8 mm, placed bilaterally between the roots of the maxillary second premolars and first molars. One quadrant received laser treatment and the contralateral quadrant served as the control | Success/stability: lower Periotest value (PTV), indicating less mobility and greater miniscrew stability. Failure/instability: higher PTV, reflecting increased mobility; clinical miniscrew failure or loss was not directly reported | Miniscrews were loaded one week after placement with approximately 200 g per side for anterior retraction. A 980-nm diode laser was applied at placement, 1 week, 1 month, and 3 months. Stability and peri-miniscrew crevicular fluid biomarkers were assessed at the same time points | Mean PTV was lower in the laser group at all time points. The difference was not significant initially or at 1 week, but laser-treated miniscrews showed significantly greater stability at 1 month (PTV −6.33 vs −5.53; P = 0.008) and 3 months (−6.67 vs −5.55; P = 0.003). IL-1β concentration was significantly lower in the laser group immediately after placement (4.89 vs 11.04; P = 0.008) and at 3 months (4.91 vs 10.00; P = 0.025). TGF-β1 levels were lower at most time points in the laser group, but no statistically significant differences were detected. The overall effects of time, laser irradiation, and their interaction on PTV were significant | Low-level laser therapy, laser wavelength and power, treatment time, miniscrew stability measured by Periotest M, IL-1β concentration, TGF-β1 concentration, time after insertion, orthodontic loading, and peri-miniscrew inflammation |
| Lee et al., 2021 | Retrospective observational study | 142 patients: 56 men and 86 women; mean age 20.7 ± 7.5 years | 280 self-drilling miniscrews, 1.5 mm in diameter and 7.0 mm long: 145 cylindrical and 135 dual-thread miniscrews. Miniscrews were placed in the buccal alveolar regions of the maxilla and mandible | Success: miniscrew maintained and clinically usable for more than 6 months. Failure: miniscrew loss or excessive mobility | Most miniscrews were loaded with 200–250 g of force in the third week after placement. The mean maintenance period was 18.8 months: 19.1 months for cylindrical miniscrews and 18.5 months for dual-thread miniscrews. Insertion and removal torque and Periotest values were recorded | Cylindrical vs dual‑thread miniscrews: overall success 82.1% (119/145) vs 84.4% (114/135), no significant difference (P=0.595). Maxilla: 79.1% vs 88.2%; mandible: 87.0% vs 78.0%. Dual‑thread had higher insertion torque in mandible (8.03 vs 6.48 Ncm; P=0.013) but no better long‑term success. Positioning success: central 91.9%, close to root 67.6%, root contact 40.0%. Risk factors: increasing age reduced failure risk (OR 0.932 per year; 95% CI 0.878–0.990; P=0.021); higher Periotest increased failure risk (OR 1.126; 95% CI 1.043–1.216; P=0.002). Root proximity greatly increased failure risk (OR 7.42 for close proximity; OR 44.13 for root contact). | Miniscrew design: cylindrical versus dual-thread, age, sex, maxilla versus mandible, side of placement, interradicular placement site, root proximity or contact, operator, insertion and removal torque, and initial Periotest value |
| Singh et al., 2021 | Prospective, randomized clinical study | Initially 40 patients aged 18–26 years; 36 patients were included in the final analysis: 20 females and 16 males. Four patients withdrew from the study | 36 titanium alloy (Ti-6Al-4V) mini-implants divided into two groups: cylindrical single-threaded mini-implants (n = 18) and cylindrical double-threaded mini-implants (n = 18). All implants were placed in the maxilla between the second premolar and first permanent molar using an implant guide | Stability: a higher MIT indicated greater primary mechanical stability during insertion, whereas MRT assessed resistance during implant removal. Failure: clinical implant loss or survival was not evaluated; the study focused on biomechanical stability parameters | Mini-implants were inserted and removed using a contra-angle handpiece and surgical engine at 30 rpm. After surgery, patients received 2% chlorhexidine mouthwash and antibiotics for 3 days. No long-term follow-up period or detailed orthodontic loading protocol was reported | Mean MIT was significantly higher for double-threaded than for single-threaded mini-implants: 8.72 ± 0.67 Ncm versus 7.22 ± 1.56 Ncm (P = 0.000642). Mean MRT was also higher in the double-threaded group: 1.78 ± 0.43 Ncm versus 1.22 ± 0.43 Ncm (P = 0.000436). In both groups, MIT was significantly higher than MRT: for single-threaded implants, 7.22 ± 1.56 versus 1.22 ± 0.43 Ncm; for double-threaded implants, 8.72 ± 0.67 versus 1.78 ± 0.43 Ncm (P < 0.00001). These findings indicate that the double-threaded design provided greater primary mechanical stability | Thread design—single versus double; maximum insertion torque; maximum removal torque; implant geometry; placement site; implant length and diameter; cortical bone thickness; insertion technique and speed; and the relationship between primary stability and removal torque |
| Lin et al., 2015 | Retrospective observational study | 136 patients; 285 miniscrew implants were initially assessed. Smokers and patients with systemic diseases or long-term medication use were excluded | Orthodontic miniscrew implants of two types: AbsoAnchor and VectorTAS. Diameters were 1.3, 1.4, or 2.0 mm; lengths were 6–7, 8, or 10–12 mm. Miniscrews were placed in the maxilla or mandible, including anterior and posterior regions, the retromolar area, and the palate | Success at T1: absence of infection, excessive mobility, dislodgement, need for immediate removal or replacement, or irreversible biological damage before orthodontic loading. Success at T2: the miniscrew remained functional under orthodontic loading without dislodgement until 12 months after insertion or until intentional removal after completing its anchorage function. Failure: dislodgement, excessive mobility, infection requiring removal, or irreversible damage to adjacent structures | Outcomes were assessed at two time points: T1, immediately before orthodontic loading; and T2, 12 months after insertion or after the miniscrew had completed its anchorage function, whichever occurred first. The mean loading time for failed miniscrews was 3.5 months; most failures occurred within the first 5 months after loading | Overall T1 success was 94.7% (270/285) and T2 success was 83.3% (214/257). Miniscrew length was the only implant-related factor consistently associated with success: 6–7 mm screws had T1 82.7% and T2 76.7%; 8 mm screws had T1 97.3% (adjusted OR vs 6–7 mm 11.88, P = 0.001) and T2 82.1%; and 10–12 mm screws had T1 98.0% (adjusted OR vs 6–7 mm 10.50, P = 0.035) and T2 93.5% (adjusted OR vs 6–7 mm 17.95; 95% CI 1.83–176.01; P = 0.013). Patients with average vertical facial pattern had higher T2 success than high-angle patients (92.6% vs 79.6%; adjusted OR 4.22; 95% CI 1.35–13.16; P = 0.013). Although Class III showed lower success in univariate analysis, this was not significant after adjustment. Miniscrew type, diameter, jaw, side, and placement site were not significantly associated with success. | Patient age, sex, sagittal and vertical skeletal malocclusion, dental malocclusion, placement side, jaw, anatomical site, miniscrew type, length, and diameter. The study also considered the influence of implant loading and clinical dislodgement, although loading force, insertion torque, insertion direction, operator experience, and exact loading protocol were not available from the records |
| Sahoo et al., 2025 | Randomized clinical trial | 32 adult orthodontic patients requiring extraction of the maxillary first premolars and anchorage reinforcement with mini-implants; 16 patients in the immediate-loading group and 16 in the delayed-loading group | 64 self-drilling orthodontic mini-implants; 1.5 mm in diameter and 10 mm in length. Two mini-implants were placed per patient on the buccal side of the maxilla, between the second premolar and first molar, in attached gingiva. Implants were inserted at an angle of 30–45° to increase cortical bone contact | Success: the mini-implant remained stable, without mobility, loss, soft-tissue ulceration, peri-implantitis, or other complication, and was able to withstand orthodontic force for more than 6 months. Failure: implant mobility, dislodgement/loss, or inability to function as orthodontic anchorage | Immediate-loading group: implants were placed after leveling and alignment and loaded immediately with 250 g using elastic chains. Delayed-loading group: implants were placed approximately 1 month before orthodontic loading. Patients were reviewed on day 1, day 7, and at 1, 3, and 6 months. Pain, discomfort, swelling, speech difficulty, chewing difficulty, mobility, implant loss, and peri-implant soft-tissue health were assessed | Overall mini-implant survival/success was 100% in both groups, with no failures and no significant difference in success between immediate and delayed loading. Pain was greatest during insertion and decreased over time in both groups. Immediate loading produced significantly greater pain, discomfort, swelling, speech difficulty, and chewing difficulty than delayed loading (reported P ≤ 0.005; the text also reports P < 0.005). Mild bleeding and ulceration were reported in the immediate-loading group. No cases of peri-implantitis or implant loss were observed | Loading protocol, immediate versus delayed force application, postoperative pain, discomfort, swelling, speech and chewing difficulties, bleeding, ulceration, implant mobility or loss, peri-implant soft-tissue condition, periodontal health, and patient sex/age. The study also standardized implant dimensions, placement site, insertion angulation, orthodontic force, and oral-hygiene instructions |
| He et al., 2023 | CBCT-based clinical and anatomical analysis of successful mini-implant insertion, with radiographic validation | 21 adult orthodontic patients with 40 successfully inserted mini-implants; an additional 54 infrazygomatic crest sites from 27 patients were assessed for validation | Orthodontic mini-implants placed in the maxillary infrazygomatic crest adjacent to the first permanent molar (U6); implant dimensions were not specified | Clinical success: no loosening, loss, root damage, sinus damage, or injury to adjacent structures until treatment completion. Radiographic success: at least 1 mm from the tooth root and more than 1 mm of total cortical bone thickness | CBCT was performed immediately after insertion. The validation analysis simulated implant placement at different heights and angles; loading timing was not reported | Successful implants were positioned 0.4 ± 2.8 mm distal and 8.2 ± 2.5 mm above the U6 CEJ, at an angle of 56.4° ± 7.7°. Mean buccal and palatal bone thicknesses were 4.1 and 7.2 mm, respectively; cortical bone thickness was 2.4 ± 0.6 mm. At the proposed position and angle, 75.9% of 54 sites met the safety criteria. Increasing implant height by 1, 2, and 3 mm increased eligibility to 83.3%, 90.7%, and 94.4%, respectively. Increasing the angle had only a limited effect | Implant position and height, insertion angle, cortical bone thickness, buccal and palatal bone thickness, root proximity, infrazygomatic crest anatomy, and potential maxillary sinus interference |
| Yildiz et al., 2024 | Prospective observational study | 16 systemically healthy, non-smoking individuals; 13 females and 3 males; mean age 21.9 ± 1.8 years | 24 self-tapping, self-drilling titanium-alloy mini-screws with a rough titanium-oxide-coated surface; 1.6 mm diameter and 7 mm length. Eighteen screws were placed in the maxilla and six in the mandible, mainly in interradicular areas | Clinical success: no early mini-screw loss or treatment-related complication. Failure: mini-screw loss or clinically relevant loss of stability; no early losses were observed in this study | Periodontal and stability assessments were performed 2 weeks after placement (T0) and at 3 months (T1). Mini-screws were removed after approximately 3.5–4 months. Stability was assessed using a torque gauge. All patients received professional cleaning and oral-hygiene instruction before screw placement | No early mini-screw loss occurred. Site-level Gingival Index decreased from 1.4 to 1.1 (p = 0.016), while site-level bleeding on probing decreased from 55.7% to 29.7% (p < 0.001). Full-mouth bleeding on probing also decreased from 38.7% to 29.6% (p = 0.003). No significant changes were found in Plaque Index, probing pocket depth, or gingival recession. Supracrestal tissue height decreased from 2.0 to 1.5 mm (p = 0.038). Radiographic analysis showed a significant reduction in bone level around the mini-screws (p = 0.008), whereas the distance between the mini-screws and adjacent roots did not change significantly. Torque-gauge values decreased from 103.3 to 63.3 units (p < 0.001) | Periodontal health, oral hygiene, plaque accumulation, gingival inflammation, bleeding on probing, probing pocket depth, gingival recession, mucosal discomfort and redness, keratinized tissue width, supracrestal tissue height, transmucosal soft-tissue thickness, alveolar bone level, screw–root distance, and torque values |
| Kalra et al., 2014 | Randomized split-mouth clinical study | 13 orthodontic patients, 10 females and 3 males, aged 14–28 years | 40 mini-implant sites: 24 in the maxilla and 16 in the mandible. Mini-implants measured 1.5 × 9 mm and were placed in the buccal interradicular space between the second premolar and first molar. Twenty sites were assigned to CBCT guidance and 20 to RVG guidance | Accurate placement: minimal deviation from the ideal height, mesiodistal entry point, implant tip, and angulation, without root contact. Failure: increased mobility requiring removal; one RVG-guided mini-implant failed during the first week of loading | Ideal placement sites were determined using preoperative CBCT. Mini-implants were inserted after orthodontic leveling and alignment. Postplacement CBCT was performed for accuracy assessment. The mini-implants were loaded for approximately 2 months | Mean deviation from the ideal height was significantly lower in the CBCT group than in the RVG group: 0.0985 mm versus 0.565 mm (p = 0.02). Differences in entry-point deviation (0.391 vs. 0.651 mm; p = 0.143), tip deviation (0.586 vs. 1.038 mm; p = 0.204), and angular deviation (4.7° vs. 6.2°; p = 0.624) were not statistically significant. Root contact occurred in 0/20 CBCT-guided implants and 3/20 RVG-guided implants (15%); all three contacts occurred in the mandibular arch. One RVG-guided implant failed because of increased mobility, although it was not in contact with a root | Imaging and guidance method, vertical placement accuracy, mesiodistal positioning, implant-tip deviation, angular deviation, root proximity, arch (maxilla versus mandible), interradicular space, cortical bone thickness, bone density, buccolingual bone thickness, and initial stability |
| Athira et al., 2023 | Retrospective CBCT-based anatomical study | 48 patients aged 18–30 years, divided into four groups: prognathic mandible (n = 12), retrognathic mandible (n = 12), horizontal growth pattern (n = 12), and vertical growth pattern (n = 12) | Mandibular buccal shelf extending from the first molar region to the area distal to the second molar. Seven sectional sites were examined: mesial, middle, and distal regions of the first molar; the interradicular region between the first and second molars; and mesial, middle, and distal regions of the second molar | Favorable site: an area providing greater cortical bone thickness, adequate insertion depth, and a reduced risk of root contact. A cortical thickness greater than 1 mm was considered clinically desirable for primary stability | CBCT measurements were obtained at 3, 5, and 7 mm from the alveolar crest. Cortical bone thickness and safe insertion depth were assessed at insertion angles of 90° and 30° relative to the long axis of the tooth. The angulation and area of the buccal shelf were also evaluated | Cortical bone thickness increased distally from the first to the second molar in all four groups. The most favorable site was distal to the mandibular second molar, approximately 7 mm from the alveolar crest, with insertion at 30°. The greatest amount of cortical bone and a favorable insertion depth were observed at 30° and 7 mm. Insertion at 90° or at 30° and 3 mm from the crest was associated with a greater risk of root proximity | Location relative to the first and second molars, distance from the alveolar crest, insertion angle (90° vs. 30°), cortical bone thickness, safe insertion depth, horizontal versus vertical growth pattern, prognathic versus retrognathic mandibular pattern, buccal shelf angulation, and buccal shelf area |
| Garg et al., 2022 | Prospective, randomized clinical study | 10 patients aged 13–35 years with bimaxillary protrusion or Class II malocclusion requiring maximum anchorage | 40 miniscrews: 20 titanium and 20 stainless steel; 1.3 mm in diameter and 8 mm in length. Miniscrews were placed between the second premolar and first molar in both the maxillary and mandibular arches | Grade 1: no mobility. Grade 2: 0–1 mm mobility. Grade 3: 1–2 mm mobility. Grade 4: >2 mm mobility or implant loss | Miniscrews were loaded one week after insertion with a 250-g force using a NiTi closed-coil spring. Stability was assessed at insertion (T0), after 1 month (T1), and after 6 months (T2). Retraction was recorded monthly for 6 months | At T0, all 40 miniscrews showed Grade 1 stability. At 1 month, Grade 2 mobility occurred in 3 titanium and 7 stainless steel miniscrews; the difference was not statistically significant (P = 0.273). At 6 months, Grade 2 mobility occurred in 0 titanium and 4 stainless steel miniscrews. Grade 3 mobility occurred in 4 implants in each group, while Grade 4 mobility occurred in 1 implant in each group. The difference remained statistically nonsignificant (P = 0.187) | Implant material, primary and long-term stability, follow-up period, dental arch, amount of anterior retraction, and anchorage maintenance during en masse retraction |
| Bollero et al., 2018 | Prospective clinical randomized split-mouth study | 15 patients: 6 males and 9 females; mean age 16.2 ± 4.6 years | 30 miniscrews: one TiA and one SS miniscrew per patient. Both types were 8 mm long and 1.5 mm in diameter. The miniscrews were placed buccally in the posterior maxillary alveolar region, mesial to the first molars | Success: the miniscrew remained stable and functional until completion of canine retraction, without infection, fracture, or clinically significant mobility. Stability was assessed using Periotest values and removal torque | Immediate loading with approximately 90–100 g of force using nickel-titanium springs. Mean functional service time was 160.8 ± 23 days | Both materials achieved a 100% success rate. Mean insertion torque was 3.2 ± 1.1 Ncm for TiA and 2.9 ± 0.9 Ncm for SS. Mean removal torque was 4.4 ± 1.3 Ncm for TiA and 5.1 ± 0.7 Ncm for SS. The differences were not statistically significant. Periotest values were 8.8 for TiA and 9.1 for SS, with no significant difference | Miniscrew material, insertion and removal torque, mechanical stability, SEM morphology, surface composition, structural damage, and evidence of osseointegration |
| Chawshli et al., 2024 | Randomized split-mouth clinical trial | 29 patients aged 18–40 years: 18 females and 11 males | 58 titanium mini-implants, 1.6 × 8 mm, placed bilaterally in the anterior maxilla between the lateral incisor and canine, in attached gingiva | Success: maintenance of clinical stability. Stability was assessed monthly using a torque-measuring device, with resistance to removal at 5 Ncm | Immediate loading with 15 g per side. Monthly follow-up for 6 months | At 6 months, stability was 0% in the 45° group and 27.6% in the 90° group (p = .004). Stability was higher in males than females: 27.3% vs. 5.6% at 6 months (p = .044) | Insertion angle, age, sex, follow-up duration, anterior maxillary site, and CBCT-guided implant placement |
| Li et al., 2022 | Retrospective anatomical CBCT study | 90 adults | No clinical TADs; simulated palatal placement between the maxillary second premolar–first molar and first–second molar | Anatomical safety based on interradicular width, palatal bone thickness, and avoidance of root or sinus contact | No clinical follow-up or loading | The safest sites were 14–16 mm from the palatal apex, generally at 60°–90°. Recommended placement: between the second premolar and first molar; approximately 60° between the first and second molars | Insertion angle, placement site, interradicular width, mucosal and bone thickness, root proximity, and sinus risk |
Reported TAD success was 79–90% in buccal alveolar sites [14, 7, 4] and ~90–96% in palatal placements [13, 6], while IZC survival over 6 months was comparable to alveolar sites [11]. However, substantial variation was also reported in other clinical settings: success was 73.7% in the maxillary tuberosity [25], 82.1–84.4% for cylindrical and dual-thread miniscrews [30], 85.1% in a prospective cohort [18], and 87.5% for mandibular buccal-shelf and interradicular miniscrews [22].
Failures clustered early: ≈70–80% occurred within 1–4 months after insertion, with a first-month peak; stability generally consolidated by 12–16 weeks [14, 5, 4]. In contrast, some studies reported later failures, particularly in mandibular buccal sites, whereas all primary failures in one prospective study occurred by day 42 [22, 26]. These findings indicate that the timing of failure may depend on anatomical site, loading protocol, operator experience, and implant design [26, 15, 25].
Root contact was the most consistent independent predictor of failure (OR ≈28) on CBCT, outweighing other factors [8]. A separate CBCT-based study similarly identified root contact as the strongest predictor of failure, with an odds ratio of 13.50; among implants contacting a root, each additional millimetre of contact length approximately doubled the failure risk [24]. Miniscrews positioned centrally between the roots showed substantially higher success than those close to or contacting a root [30].
In one cohort, greater distance from the alveolar crest was associated with lower failure odds, indicating the benefit of denser, thicker bone away from the crest [10]. CBCT-based anatomical studies further indicated that greater cortical bone thickness, adequate insertion depth, and a greater distance from the palatal suture may improve primary stability [23]. In the mandibular buccal shelf, the most favourable site was distal to the second molar, approximately 7 mm from the alveolar crest, with an insertion angle of about 30° [37].
In the palate, smaller pilot holes (60% of screw diameter; e.g., 1.2 mm for 2.0-mm screws) increased primary stability and success, with effective torque bands around 20–30 N·cm [13]. Longer miniscrews (10–12 mm) improved alveolar survival compared with 8–9 mm [7], whereas in IZC, longer length raised primary stability without survival benefit and increased sinus penetration and early pain, supporting shorter screws for safety [11]. Surface treatment enhanced removal torque (secondary stability) but did not clearly reduce failures within limited samples [3]. Similarly, double-threaded miniscrews produced higher insertion and removal torque than single-threaded designs [31], although differences in long-term clinical success between cylindrical and dual-thread miniscrews were not statistically significant [30]. Titanium and stainless-steel miniscrews also demonstrated comparable clinical stability and success, suggesting that implant material alone is not a dominant determinant of outcome [38, 39].
Very early loading (≤1 week) increased failure in alveolar cohorts; after 2–4 weeks, loading effects were inconsistent [5, 7]. Nevertheless, a randomized trial found 100% success with both immediate and delayed loading when implant dimensions, insertion angle, force magnitude, placement site, and hygiene instructions were standardized [33]. Immediate loading was associated with more pain, swelling, speech difficulty, chewing difficulty, mild bleeding, and ulceration, but not with increased implant loss [33].
Associations between vertical skeletal pattern and TAD outcome were inconsistent. [15] reported more early failures at lower FMA values, whereas [8] observed higher FMA values in the failure group on bivariate analysis, with no independent effect in multivariable analysis. [14] also reported an association between vertical skeletal pattern and success. These findings should therefore be regarded as heterogeneous and site-specific rather than as evidence of a uniform effect.
In 467 buccal molar TADs, overall failure was 15.6% and was higher in the mandible, males, and patients aged ≥30 years. Early failure increased as FMA decreased, reaching 19.0% in patients with FMA ≤20° [15]. Low FMA may represent a marker of increased early mechanical risk in some posterior buccal cohorts, although the anatomical and temporal specificity of this association remains uncertain. The potential role of predrilling and torque-controlled insertion in such cases remains to be tested prospectively.
Age and sex were not consistently associated with failure: some studies reported higher failure in males, older patients, or patients aged over 45 years [15, 17], whereas other cohorts found no significant demographic effects [18, 8]. In one large cohort, operator experience was the only significant predictor, with implants placed by less-experienced operators showing approximately fourfold higher odds of failure [25].
Among 121 immediately loaded TADs, overall success was 89.3%. Success was higher in vestibular than palatal sites, although the difference was not statistically significant. In 958 TADs used for maxillary distalization, the failure rate was only 1.5%. Failures were more frequent in palatal sites and in patients older than 45 years. A standardized protocol involving pilot drilling, controlled insertion, and immediate loading may have reduced technical variability [17]. The particularly low failure rate in this cohort suggests that immediate loading may be predictable when placement is performed using a standardized protocol with controlled angulation, pilot drilling, and a defined loading force [17].
In a prospective study of 47 TADs, success was 85.1%. Early mobility occurred in 14.8% of implants, most often during the second week. No significant association was found with age, sex, jaw, placement site, mucosal type, or screw dimensions [18]. Peri-implant inflammation was associated with early instability. During the first week, success decreased from 92.9% without inflammation to 33.3% with moderate inflammation; during the second week, it decreased from 96.3% to 63.6%, respectively [18]. Pain was common shortly after insertion but decreased over time and was not associated with TAD success. Therefore, pain alone should not be considered a predictor of failure [18].
Immediate loading may be successful when placement, force, and case selection are carefully controlled. However, its effects depend on implant design, bone quality, insertion technique, and soft-tissue condition [16, 18]. A randomized trial of low-level diode laser therapy reported lower Periotest values and greater stability at 1 and 3 months, together with reduced IL-1β concentrations, suggesting that modulation of peri-implant inflammation may improve early stability [29]. However, the study assessed stability rather than definitive implant survival and included a small final sample.
Vertical skeletal pattern and other patient-related characteristics correlated with lower success in posterior buccal placements in some analyses [14, 15], but the association was mainly observed during the first 120 days in patients with lower FMA values in the Yoshida cohort [15], whereas the Yamaguchi study reported higher FMA values in the failure group on bivariate analysis, without an independent effect in multivariable analysis [8]. In the prospective study by Omer, age and sex were not significant predictors of success [18].
Site-related soft-tissue context influenced complications: buccal fold showed more mobility/inflammation than palate; intrusion mechanics had higher loosening than extrusion [6]. Keratinized mucosa showed numerically higher success, although mucosal type was not statistically significant [18]. IZC sinus penetration was common radiographically yet usually asymptomatic; deeper penetration may elevate risk [11]. In the IZC trial, sinus penetration occurred in more than half of the implants and was numerically more frequent with 14-mm than 12-mm screws, while longer screws also produced greater pain on the first postoperative day. These findings support choosing the shortest screw that provides adequate mechanical stability [11].
Biological correlates of failure included elevated peri-implant IL-6 [9] and dysbiotic microbiota enriched in periodontal pathogens, notably Prevotella intermedia [1]. Peri-implant inflammation may also serve as an early clinical marker of failure [18]. The microbiome study further showed distinct microbial profiles between failed and successful TADs, with enrichment of Prevotella intermedia, Eikenella corrodens, Parvimonas species, and other periodontal pathogens in failed implants. Prevotella intermedia demonstrated promising discriminatory ability for failure, although these findings remain preliminary [1].
Insertion technique and imaging guidance may also influence the risk of anatomical complications. CBCT-guided placement produced significantly lower vertical deviation than two-dimensional radiography combined with a surgical guide, and root contact occurred only in the radiography group in one randomized study [36]. However, the differences in mesiodistal, transverse, and angular deviations were not statistically significant. A separate randomized study found that a 3D-printed surgical guide produced lower platform deviation than real-time navigation or freehand placement, although no immediate clinical complications were observed [12].
For anterior maxillary intrusion, insertion angle appeared particularly important: 90° placement showed greater six-month stability than 45° placement, although the absolute stability rates were low in both groups. Male patients also demonstrated higher six-month stability than female patients in that study; because of the small sample and high overall failure rate, these findings should be interpreted cautiously [40].
Loading time should therefore be interpreted together with implant design, insertion technique, force magnitude, bone quality, and soft-tissue health [16, 18]. The available evidence also suggests that a healing period may be particularly important in conventional alveolar sites, whereas immediate loading can be acceptable in carefully selected cases with high primary stability [5, 33, 17]. Immediate loading should therefore be regarded as a protocol-dependent clinical option rather than a universally equivalent alternative to delayed loading.
Methodological heterogeneity, nonstandard success/failure definitions, variable follow-up, and incomplete reporting of torque, loading, and tissue conditions limit direct comparison and pooling across studies [14, 7, 8]. Additional limitations arise from the use of surrogate outcomes, such as insertion or removal torque, Periotest values, and radiographic safety criteria, which do not always correspond to long-term clinical survival [31, 29, 28]. Several studies also included small samples, single-centre populations, or multiple implants per patient without fully accounting for clustering. Consequently, the observed associations should be considered hypothesis-generating rather than definitive evidence of causality.
The included studies suggest that TAD outcomes vary across anatomical sites and treatment protocols. Failures frequently occurred during the first few months after insertion, although their timing differed between studies [4,5,14]. Root contact was associated with failure in several studies [8,24,30]. These findings identify anatomical relationships as an important consideration when interpreting TAD outcomes, but they do not establish a single placement protocol applicable to all sites.
The findings also distinguish initial mechanical stability from longer-term clinical success. Longer screws were associated with better survival in some alveolar sites, whereas greater length in the infrazygomatic crest improved primary stability without a clear survival benefit and was associated with sinus penetration and early pain [7,11]. Similarly, findings on palatal pilot-hole preparation cannot automatically be extended to other anatomical sites [13]. Evidence on loading time was inconsistent: early loading was associated with failure in some cohorts, while a randomized trial reported no failures with either immediate or delayed loading [5,7,33]. These differences limit conclusions about the independent effect of loading time.
Associations with skeletal pattern and demographic characteristics were also inconsistent. Low FMA values were associated with early failure in the Yoshida cohort, whereas the association with higher FMA values in the Yamaguchi study did not remain independent in multivariable analysis [8,15]. Findings on age and sex likewise differed between studies [8,15,17,18]. These variables therefore cannot be treated as uniform predictors of failure across clinical settings.
Biological findings require similarly cautious interpretation. In one prospective study, peri-implant inflammation was associated with early instability, whereas postoperative pain decreased over time and was not associated with success [18]. Differences in inflammatory markers and microbial composition were also reported between successful and failed devices [1,9]. These associations do not establish whether the biological changes preceded instability or developed alongside it, and their value for predicting failure remains uncertain.
Previous reviews provide context for the present findings. Schätzle et al. reported differences in failure rates between TAD types [19], while Ramírez-Ossa et al. identified factors related to the device, patient, procedure and orthodontic treatment that were associated with TAD outcomes [20]. Reynders et al. specifically examined the relationship between insertion torque and mini-implant success and highlighted the need for more objective stability measurements [21].
This review considers reported complications alongside TAD success and failure outcomes, bringing together findings on anatomical, mechanical, inflammatory and microbiological factors. The structured presentation of study characteristics, outcome definitions and loading protocols provides context for interpreting differences between studies. Independent screening and data extraction by two reviewers, with disagreements resolved through discussion, were used to reduce errors in study selection and data collection.
The search was limited to PubMed, Europe PMC and BASE, and inclusion was restricted to publications in English with freely accessible full texts. Relevant studies available through other sources, in other languages or behind paywalls may therefore have been omitted. The full-text restriction reflected the lack of institutional access to subscription-only articles and may have introduced availability bias.
The included studies differed in design, sample size, anatomical site, device characteristics, loading protocol, follow-up duration and outcome definitions. These differences limited direct comparison of reported success, failure and complication rates. Retrospective designs and small samples in some studies also limit confidence in the reported associations.
Potentially relevant variables, including bone quality, oral hygiene, insertion torque, loading force and soft-tissue condition, were not consistently reported. Some studies assessed mechanical or radiographic measures rather than clinical survival or complications, and these outcomes should not be treated as equivalent. Associations identified in observational studies do not establish causality.
This review provides a descriptive synthesis without pooled estimates of success or failure. Heterogeneity among the included studies and limitations of the available comparisons prevent firm conclusions about the relative performance of different devices, placement techniques or loading protocols.
Reported TAD success, failure and complication outcomes varied across anatomical sites and treatment protocols. Failures frequently occurred during the first few months after insertion. Root contact was associated with failure in several studies, while peri-implant inflammation was associated with early instability in one prospective study. Postoperative pain and sinus penetration were also reported. The predictive value of inflammatory and microbiological findings remains uncertain.
Findings concerning device dimensions, insertion procedures, loading time and patient characteristics were inconsistent or specific to particular clinical settings. The available evidence does not support a single protocol for TAD selection, placement or loading.
Differences in study design, follow-up duration and outcome definitions, together with incomplete reporting of relevant variables, limited comparisons between studies. Future prospective studies should distinguish primary stability from clinical success throughout the intended period of orthodontic loading and use standardized outcome definitions and reporting methods.
Conceptualization: Tomasz Horodniczy
Methodology: Tomasz Horodniczy, Justyna Polko, Izabela Migdał
Data collection: Katarzyna Raczek, Weronika Kwaśnica
Analysis: Zuzanna Jeziorska, Bartosz Szymajda
Writing: Jakub Artur Czapkiewicz, Zuzanna Galicka, Maja Witek
Supervision: Tomasz Horodniczy
This scoping review was based on previously published studies and did not require ethical approval.
The authors declare no conflicts of interest.
This research received no external funding.
Artificial intelligence tools were used to assist with language editing and formatting. The authors take full responsibility for the content of the manuscript.